LCMS-2020 - Značajke

High-Performance Liquid Chromatograph Mass Spectrometer

UFswitching

High-speed polarity switching between positive/negative ion modes

To detect both positive and negative ions, analysis is performed while switching between the positive and negative ionization modes. The LCMS-2020 adopts patented high-voltage power supply technology (Patent: US7855355) to achieve an ultra-fast polarity switching time of just 15 msec.

UFsensitivity

Superior sensitivity, especially for ultra high-speed analysis

Superior sensitivity, repeatability, and linearity are achieved, thanks to the newly developed Qarray® optics.

UFscanning

Fast scanning speeds

In MS scanning, a range of mass to charge ratios are determined in successive 0.1 m/z steps. This is controlled by applying a rapidly changing voltage to the quadrupole (quadrupole rods). By adopting Shimadzu's proprietary scanning technology (Patent: US8188426), high ion transmission is achieved without sacrificing sensitivity or resolution, even during high-speed scans up to 15,000 u/sec. Very narrow UFLC peaks can now be properly detected in scan mode acquisition.

{"title":"Downloads","description":"Preuzmite najnoviju bro\u0161uru.","source":"product","key":3151,"max":30,"filter_types":["brochures"],"link_title":"Pogledajte ostala preuzimanja","link_url":"","pdf_links":[]}